Frequently Asked Questions
Click on your product below to see a list of corresponding FAQs.
The GREF subtracts the value you place there from the actual reading that you have.
The GREF subtracts the value you place there from the actual reading that you have.
The driver will not work on multiple USB ports. The first thing to try is to plug the USB adapter cable into the correct USB port, plug the other end into the CG-5, and then turn on the CG-5 itself. The connection should be recognized by the Windows as a USB device. If this does not happen , please contact Customer Service for more information on troubleshooting the SCUTIL driver program.
Please check the fuse of the console in the battery compartment on the back of the ENVI/ENVIPRO with an ohm meter. It should be a few ohms. Replace if needed.
Low battery; not enough current available for the injection sequence.
If the temperature is below -15C then activate the display heater by pressing the SETUP key and then changing the HEATER to yes.
The sensor is low in Hexane. Refill the sensor to a total weight (sensor and Hexane) of 960-980 grams.
Make sure that the sensor orientation is correct with the sensor N pointed to magnetic North.
You may be in a noisy environment, try a remote location free from power lines and buildings.
The sensor cable is broken, replace the cable.
Check the connection of the GPS to the ENVIPRO console.
In some instances the canopy in the survey area may block the signal and the GPS will not lock.
If you are using the cable system, please make sure that the takeouts of the cable are correctly attached to the electrodes and are in the proper order. If this is correct, cold boot the system, re-enter your survey data and try once more. If the situation does not improve please contact Customer Service.
The first thing to do is to cold boot the system and renter your survey data. Once this is done reinitialize the cable detection as per the manual on page 2-7.
If you still have the error, disassemble the middle module of the SARIS from the main SARIS console. (between the battery and the top console).
Loosen and remove the 4 screws on the module top and remove the top plate to access the two fuses in the module. Check both fuses for continuity by using an ohm meter. They should be a few ohms. Replace if needed.
If you have a cable interface module, make sure that the ribbon cable to the SARIS main console is connected securely.
If the WINDOWS does not recognize the SARIS console as a USB device, reinstall the SCUTIL as per page C-2 of the manual. Please mark on the computer the USB port used as the SCUTIL is not plug_and_play compatible and can be installed on one USB port only.
Can’t find the answer to your question?
Contact our Support team for further assistance.